Turbidimetric Determination of Fecal Calprotectin Using Two Table Top Chemistry Analyzers: Mindray BS-200E and Cobas® c111

BÜHLMANN fCAL® turbo:

Mandic-Havelka, A. Nilsen, T. et. al. Turbidimetric Determination of Fecal Calprotectin Using Two Table Top Chemistry Analyzers: Mindray BS-200E and Cobas® c111.  Clin Lab.  2017 May 1;63(5):907-913. doi: 10.7754/Clin.Lab.2016.161032.  PMID: 28627817

Highlights from this Publication

“The assay was linear in the range between 20 and 1,900 µg/g with a limit of quantification around 20 µg/g on both instruments. The total coefficient of variation was < 7% in the range between 50 and 1,300 µg/g on both instruments. No antigen excess hook effect was observed up to 18,000 µg/g on the Mindray BS-200E and up to 20,000 µg/g on cobas® c111. The BÜHLMANN fCAL® turbo assay showed a high correlation with the BÜHLMANN fCAL® ELISA.”…

“No carry over was detected either on the cobas  c111 nor on the Mindray BS-200E.”

BÜHLMANN fCAL® turbo (K190784): FDA 510(k) cleared. For in vitro Diagnostic Use.